Guidance and manufacturer's declaration – electromagnetic immunity
The SmartLite Pro® System is intended for use in the electromagnetic environment specified below. The customer or the
user of the SmartLite Pro® System should assure that it is used in such an environment.
IMMUNITY Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is a ected by
absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
a
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the SmartLite Pro®
System is used exceeds the applicable RF compliance level above, the SmartLite Pro® System should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the SmartLite Pro® System.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
22
IEC 60601
Test level
3 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2,7 GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to any part of
the SmartLite Pro® System, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = [ 3,5 ] √P
3
d = [ 3,5 ] √P
3
d = [ 7 ] √P
3 Vrms
3
Where P is the maximum output power rating
10 V/m
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
80 MHz to 800 MHz
800 MHz to 2,7 GHz
b
a