Descargar Imprimir esta página

Dentsply Sirona SmartLite Pro EndoActivator Manual De Instrucciones página 16

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 82
Cleaning and
Cleaning
Disinfection:
The handpiece, attachment and charging base have to be manually cleaned.
Manual
1. Discard used gloves according to local regulations.
2. Disinfect hands with an appropriate bactericidal, virucidal and fungicidal hand
3. Use a new clean pair of examination gloves.
4. Disassemble attachment from handpiece. Use separate wipes for attachment and
5. Scrub handpiece, attachment and charging base with an impregnated wipe or
6. Special Note: Use care when cleaning the mating surfaces of the attachment and
7. Remove cleaning solution residue with a damp cloth. Use tap water to dampen cloth.
8. Allow the devices to air dry for at least 5 minutes.
Cleaning and
Disinfection
Disinfection:
1. After cleaning, wipe all device surfaces with a new single-use cloth in combination with
Manual
2. Ensure that the device stays wet for the entire contact time specified by wrapping
3. Pay special attention to seams, areas around buttons, window and crevices.
4. Use fresh wipes to disinfect the attachment o-ring area, handpiece mating cavity
5. Wipe the devices with a sterile, clean, lint-free cloth that is well dampened with
6. Wipe device with a fourth dry, sterile lint-free cloth to remove all fluid.
7. Allow the devices to air dry for at least 5 minutes.
Drying
Wipe the devices dry with a sterile, clean, lint-free cloth. Allow the components to fully air
dry before storage.
Maintenance,
• Visually inspect to ensure that all contamination has been removed.
Inspection and
• Visually inspect power supply and cord for damage.
Testing
• Components that are damaged, worn, or distorted such as the o-rings should be
discarded and replaced.
• See maintenance section below for additional recommended maintenance and testing.
16
disinfectant solution according to local regulations. Use according to disinfectant
solution manufacturer's Instructions for Use.
handpiece.
disposable towel soaked with an alcohol-based, tuberculocidal, quaternary ammonium
solution with a label claim for cleaning, approved according to local regulations and
use according to cleaning solution manufacturer's Instructions for Use until it is free of
visible residues.
handpiece. Use only a moist impregnated towel.
• For the attachment: Vigorously scrub the area near the o-ring with a fresh wipe.
Ensure fluid covers o-ring and surrounding crevices. When cleaning the mating
surface, ensure that cleaning agent only contacts the sides that fit within the
handpiece (with o-ring). Avoid applying cleaning agent to the electrical contacts on
the bottom of the attachment.
• For the handpiece mating cavity: Use a fresh wipe to clean mating groove directly
below the surface. Use care to ensure cleaning agent is applied only to the top of
the cavity interior. Ensure only minimal cleaning agent enters the cavity that houses
the electrical pins. Do not allow fluid to pool in the cavity around the contact pins.
Immediately absorb excess fluid with a dry disposable towel.
• For the battery and handpiece mating seam: Use a fresh wipe to clean
mating groove.
• Remove all visible soil, ensuring fluid penetrates all crevices. Use fresh wipes to rub
fluid into the crevices. Do not allow solution to penetrate the casing. Discard used
wipes. Additional wipes may be used.
an alcohol-based, tuberculocidal, quaternary ammonium solution, approved according
to local regulations and use according to disinfectant solution manufacturer`s
Instruction for Use. Use a separate wipe for attachment and handpiece. Ensure direct
contact of device and disinfectant by pressing the wet wipes on the device after half of
the required contact time.
wipes around device. Use additional wipes as needed.
and battery/handpiece mating seam for the entire contact time. Use care to ensure
cleaning agent is applied only to the top of the cavity interior. Ensure only minimal
cleaning agent enters the cavity that houses the electrical pins. Immediately absorb
excess fluid with a dry disposable towel.
deionized water for 30 seconds to remove all disinfecting agent. Pay special attention
to all seams, especially around the attachment/handpiece junction. Ensure cloth is
damp with deionized water for the entire 30 seconds. Discard used cloth and repeat
rinsing with a new, second dampened cloth for 30 seconds. Discard second cloth and
rinse with a new, third dampened cloth for a final 30 seconds.

Publicidad

loading