Table 7: Guidance and manufacturer's declaration – electromagnetic immunity
The Heli-FX Applier is intended for use in the electromagnetic environment specified below.
The customer or the user of the Heli-FX Applier should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
21
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5% U
(> 95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% (30% dip in U
)
T
for 25 cycles
< 5% (> 95% dip in U
)
T
for 5 sec
Note: - U
is the a.c. mains voltage prior to
T
application of the test level.
3 A/m
Compliance
Electromagnetic environment-
level
guidance
± 6 kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
± 8 kV air
the relative humidity should be at least 30%.
Not Applicable
Not Applicable
Not Applicable
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.