Figure 6. Insert Drive Cable and
Foot Pedal into Control Unit
CAUTION:
bend the flexible drive cable in a diameter of
less than 8 inches (20 centimeters). A sharply
bent or kinked drive cable may cause the
control unit to overheat and stop. During a
procedure, a minimum distance of 5   feet
(1.5 meters) should be maintained between
the control unit and the tissue removal
device to allow the drive cable to hang in
a large arc with no bends, loops, or kinks.
5 Non-sterile person attaches the tissue removal
device vacuum tubing to the corresponding
connection on the tissue trap of the collecton
canister as shown in Figure 7.
Figure 7. Attach Vacuum Tube to
Collection Canister
DO NOT attempt to sharply
MyoSure Tissue Removal System IFU
Set-up: Fluent Fluid Management System
The tissue removal device is EtO sterilized. Verify that
the tissue removal device is sterile prior to use.
Do not use if the package is opened or damaged.
Discard all opened, unused devices.
CAUTION:
is intended for single use only. DO NOT
RE-STERILIZE. DO NOT REUSE. Discard
tissue removal device after use. Dispose of
the tissue removal device and packaging
according to your facility's policies and
procedures
concerning
materials and sharps waste.
WARNING: DANGER:
if used in the presence of flammable
anesthetics.
Review the System Configuration Diagram in Figure
1.
8. for set-up outline.
Figure 8. System Configuration for
Fluent Fluid Management System (See Table 1)
Refer to the Fluent Fluid Management System
2.
Operator's Manual for instructions on how to set up
the Fluent Fluid Management System.
Connect the foot pedal tube to the connector on
3.
the front of the Fluent Fluid Management System
Console.
CONNECTING THE TISSUE REMOVAL DEVICE TO
THE FLUENT FLUID MANAGEMENT SYSTEM
Remove the tissue removal device from the sterile
1.
package.
Sterile person hands the flexible drive cable and
2.
suction tube to the non-sterile person.
Non-sterile person inserts the flexible cable into
3.
the corresponding adapter on the Fluent Fluid
Management System as shown in Figure 9.
The tissue removal device
biohazardous
Risk of explosion