Atrial capture control
Parameter
Range of values
Atrial capture control
ATM (monitoring only)
ON; OFF
Min. amplitude
0.5 ... (0.1) ... 4.8 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
Safety margin
0.5 ... (0.1) ... 1.2 V
Search type
Interval; time of day
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
Time of day
00:00 ... (10 min) ... 23:50
Ventricular capture control
Parameter
Range of values
Ventricular capture control
ATM (monitoring only)
ON; OFF
Min. amplitude
0.7 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
Safety margin
0.3 ... (0.1) ... 1.2 V
Search type
Interval; time of day
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
Time of day
00:00 ... (10 min) ... 23:50
Lead configuration
Parameter
Pacing polarity A
Pacing polarity V
Sensing polarity A
Sensing polarity V
Standard
ON
1.0 V
3.0 V
1.0 V
Time of day
24 h
02:00 hh:mm
Standard
ON
0.7 V
3.0 V
0.5 V
Time of day
24 h
02:00 hh:mm
Range of values
Standard
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
IEGM recordings
Parameter
IEGM recordings
Types of IEGM recordings
IEGM recording prior
to event
IEGM signal
Rates for statistics
Parameter
High atrial rate (HAR)
High ventricular rate (HVR)
High frequency counter
Pacing and Sensing: Single-chamber
Pulse amplitude and pulse width
Parameter
Pulse amplitude
Pulse width
Sensitivity
Parameter
Sensitivity
16
Range of values
12 (quantity); each max. 10 s
High atrial rate (HAR)
Mode switching
High ventricular rate (HVR)
0; 25; 50; 75; 100%
Filtered; unfiltered
Range of values
100 ... (5) ... 250 bpm
600; 572 ... 245; 240 ms
150 ... (5) ... 200 bpm
400; 387 ... 308; 300 ms
4; 8; 12; 16
Range of values
0.2 ... (0.1) ... 6.0 ... (0.5) ... 7.5 V
0.1 ... (0.1) ... 0.5; 0.75; 1.0; 1.25; 1.5 ms
Range of values
AUTO
0.5 ... (0.5) ... 7.5 mV
Standard
HAR
ON
OFF
75%
Filtered
Standard
200 bpm
300 ms
180 bpm
333 ms
8
Standard
3.0 V
0.4 ms
Standard
AUTO