Tolerances of Parameter Values
Dual-chamber
Parameter
Range of values
Basic rate
30 ... 100 bpm
102 ... 195 bpm
200 bpm
Basic interval
1000 ms
Magnet rate
90 bpm
Magnet interval
664 ms
AV delay
15 ... 350 ms
A/V pulse amplitude
0.2 V
0.3 ... 7.5 V
A/V pulse duration
0.1 ... 0.4 ms
0.5 ... 1.0 ms
1.25 ... 1.5 ms
Sensitivity A
0.1 mV
45502-2-1 triangle pulse
0.2 ... 0.5 mV
0.6 ... 7.5 mV
Sensitivity V
0.5 mV
45502-2-1 triangle pulse
1.0 ... 7.5 mV
Refractory period A
300 ... 775 ms
Refractory period V
200 ... 500 ms
PVARP
175 ... 600 ms
PVARP after PVC
325 ... 600 ms
Max. activity rate
80 ... 100 bpm
105 ... 180 bpm
Upper rate
90 ... 190 bpm
200 bpm
High rate protection
200 bpm
en • English
Parameter
Lead impedance
Tolerance
+/-1.5 bpm
Single-chamber
+/-2.0 bpm
Parameter
+0.0/-3.0 bpm
Basic rate
+/-20 ms
+/-1.5 bpm
+/-20 ms
Basic interval
+20/-5 ms
Magnet rate
+/-0.10 V
Magnet interval
+20/-25%
Pulse amplitude
+/-0.04 ms
+/-0.10 ms
Pulse width
+/-0.15 ms
+0.2/-0.1 mV
+/-0.1 mV
Sensitivity
+/-20%
45502-2-1 triangle pulse
+/-50%
Refractory period
+/-20%
Max. activity rate
+10/-30 ms
+10/-30 ms
High rate protection
+10/-30 ms
Lead impedance
+10/-30 ms
+/-1.5 bpm
+/-2.0 bpm
+/-2.0 bpm
+0/-2.0 bpm
+20/-0 bpm
19
Range of values
Tolerance
100 ... 200 ohm
+/-50 ohm
201 ... 2500 ohm
+/-25%
Range of values
Tolerance
30 ... 100 bpm
+/-1.5 bpm
102 ... 195 bpm
+/-2.0 bpm
200 bpm
+0.0/-3.0 bpm
1000 ms
+/-20 ms
90 bpm
+/-1.5 bpm
664 ms
+/-20 ms
0.2 V
+/-0.10 V
0.3 ... 7.5 V
+20/-25%
0.1 ... 0.4 ms
+/-0.04 ms
0.5 ... 1.0 ms
+/-0.10 ms
1.25 ... 1.5 ms
+/-0.15 ms
0.5 mV
+/-50%
1.0 ... 7.5 mV
+/-20%
200 ... 500 ms
+10/-30 ms
80 ... 100 bpm
+/-1.5 bpm
105 ... 180 bpm
+/-2.0 bpm
200 bpm
+20/-0 bpm
100 ... 200 ohm
+/-50 ohm
201 ... 2500 ohm
+/-25 %