Small Massage Ball
Large Massage Ball
Rolling Pin
* Based on worst-case parameters from device with highest potential risk within the device family (40 W)
** NOHD of the system is based on this value, since it is the worse-case hazard distance
Warning: To avoid patient discomfort, keep the treatment head perpendicular to the skin surface when performing on-contact
treatments with the Large massage ball and Rolling Pin.
3.6.
Fiber Management
The console has an easy-to-use Fiber Manager and Hand Piece Clip that provides storage for the beam
delivery assembly while not in use. While paused, or between treatments, the user may store the hand
piece in the Hand Piece Clip (Figure 17).
When wrapping the hand piece fiber around the fiber manager spool, the first wrap should be fed into the
Fiber Slot so the plastic Bifurcator, which couples the finger switch and fiber optic cables, snaps into the
slot (Figure 18). This feature provides strain relief for the fiber optic connection during transport or use.
Continue with five more wraps (for a total of six) around the spool. When wrapped tightly, there should
be enough length for the hand piece to feed into the Hand Piece Clip.
3.7.
Laser Safety Eyewear
Warning:
ALWAYS wear the protective eyewear supplied with this laser system when operating the device.
•
All personnel who are in the laser treatment controlled area MUST wear laser safety eyewear.
•
DO NOT remove protective eyewear until the operator returns the laser device to Standby mode.
•
•
If you suspect injury, such as direct exposure to the eyes, seek medical attention immediately.
Additional protective eyewear can be purchased from company webstore.
3.8.
Medical Grade Power Cord
Caution: Do not use any power cord other than the one provided with the system.
If a replacement is needed, additional power cords can be purchased from the company webstore.
4. Operation
4.1.
Safety Precautions
Anyone in the laser treatment controlled area during laser operation (including the operator, all assistants, and the patient) MUST wear the
protective eyewear supplied with this laser system.
Recommendation: DO NOT allow any nonessential personnel into the laser treatment controlled area during any laser procedure.
RSR-000885-000 (6)
0.226
0.057
0.065
Figure 18
1.8
8.0**
6.7
Figure 17
Page 8 of 20