Electromagnetic Immunity Guidance
For electromagnetic immunity, essential performance is: no knife movement without activation of the firing trigger.
Guidance and manufacturer's declaration—electromagnetic immunity
The ECHELON FLEX™ Powered Plus Articulating Endoscopic Linear Cutters are intended for use in the electromagnetic environment
specified below. The customer or the user of the ECHELON FLEX™ Powered Plus Articulating Endoscopic Linear Cutters should ensure that
they are used in such an environment.
IMMUNITY test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for input/output
IEC 61000-4-4
lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
<5 % U
(>95 % dip in U
for 0.5 cycles
Voltage dips, short
40 % U
interruptions and voltage
(60 % dip in U
variations on power
for 5 cycles
supply input lines
70 % U
(30 % dip in U
IEC 61000-4-11
for 25 cycles
<5 % U
(>95 % dip in U
for 5 s
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
Compliance level
test level
± 6 kV contact
± 8 kV air
Not applicable
Not applicable
T
)
T
T
)
T
Not applicable
T
)
T
T
)
T
3 A/m
3 A/m
8
Electromagnetic environment —
guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Not applicable
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.