Status messages
Status messages are superimposed in the right upper area of the screen if
there are changes to device status.
Status message on screen
2D Camera Head
Connect camera head to control unit.
Connect compatible camera head to control
unit.
Light source on
Light source off
Check light guide connection.
Warning! Symbols of activated image effects
will not be displayed.
White balance successful
White balance failed
Camera head buttons out of service due to
magnetic field.
2.1.2
Scope of Supply
Art. no.
Designation
PV480
Full HD CMOS camera control unit
PV437
DVI cable, length 3.0 m
OP942
MIS-Bus cable length 0.75 m
TA014621
Instructions for use
TPO003-121-01
Open source license brochure
6
Status/Status Change
A compatible camera head is attached to the
camera control unit and recognized.
No camera head is connected to the camera
control unit.
No compatible camera head is connected to
the camera control unit.
The light source is switched on or off.
The light lead is not plugged at all or not cor-
rectly to the light source.
The display of the status symbols in the user
menu is deactivated before restarting the
camera control unit. The status symbols are
only displayed for a short time during activa-
tion/deactivation of the relevant function.
After activation of the white balance there is a
display of whether this has been done success-
fully.
An external magnetic field disrupts the func-
tion of the buttons on the camera head. Oper-
ation using the buttons on the camera control
unit is still possible.
Remedy
Check the plug connection between the camera head
and the camera control unit.
If the message appears repeatedly: contact your
B. Braun/Aesculap partner or Aesculap Technical Ser-
vice.
Check compatibility between camera head and camera
control unit.
If the message appears repeatedly: contact your
B. Braun/Aesculap partner or Aesculap Technical Ser-
vice.
Check the plug connection between the light lead and
the light source.
If the message appears repeatedly: contact your
B. Braun/Aesculap partner or Aesculap Technical Ser-
vice.
Reactivate status symbols in the menu.
Restart the device.
If the message appears repeatedly: contact your
B. Braun/Aesculap partner or Aesculap Technical Ser-
vice.